Reprocessing instructions | Perseus A500 21
Reprocessing
3.5.2 Surface disinfection with cleaning
Components:
Device surfaces
Trolley and holders
Prerequisites:
The surface disinfectant has been prepared in accordance with the
manufacturer's instructions.
The manufacturer's instructions, e.g., regarding shelf life or application
conditions, are observed.
An uncontaminated, lint-free cloth soaked in surface disinfectant is used for the
cleaning surface disinfection.
Cleaning
1. Wipe off obvious soiling with a disposable cloth soaked in surface disinfectant.
Dispose of the cloth.
2. Wipe all surfaces. After that, there must no longer be any soiling visible.
Components Surface disin-
fectant
Manufacturer Concentration Contact
time
Device surfaces
Dismozon plus Bode Chemie 1.6 % 15 min
Trolley and
holders
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Electric shock
Device malfunctions
Ensure that no liquid penetrates the device.
CAUTION
Risk of contaminating the device
If the water trap is removed, fluid may get into the main device while it is being
disinfected.
Leave the water trap fitted during surface disinfection.
Do not reprocess the inlets to the gas measurement system or the water traps
with elevated pressure via spraying or compressed air.

Other manuals for Dräger Perseus A500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Perseus A500 and is the answer not in the manual?

Dräger Perseus A500 Specifications

General IconGeneral
BrandDräger
ModelPerseus A500
CategoryMedical Equipment
LanguageEnglish

Related product manuals