Reprocessing instructions | Perseus A500 43
Reprocessing
3.7.2 Preparation before next use of device
3.7.2.1 Assembling and fitting patient-specific accessories and consumables
● Complete the device with the following accessories:
– Breathing hoses
– Filter
– Breathing bag hose with breathing bag
– Water trap
– Sample line
Fitting the breathing circuit and the filters
WARNING
Risk due to particles and dust
In order to protect the patient from particles and dust, a filter must be
used between the inspiratory limb of the breathing system and the
patient.
Use a filter at the Y-piece or at the inspiratory port.
WARNING
Risk of underpressure in the lungs
In the following cases, the sample gas flow can immediately lead to
underpressure in the lungs, particularly with neonates or pediatric patients:
– Filters or hoses are obstructed.
– The sample gas is extracted between the patient and an obstructed
component.
Ensure the following when ventilating pediatric patients and neonates:
When a fine pored filter is used close to the patient:
– Do not connect the sample line between the tube and the filter, but
instead directly to the filter or Y-piece.