42 Reprocessing instructions | Perseus A500
Reprocessing
3.7.1.5 Fitting the reusable CO2 absorber
If there is no CLIC absorber being used, use the reusable CO
2 absorber.
1. Push the absorber insert (A) into the absorber container (B).
2. Fill the CO
2 absorber with fresh soda lime to the upper mark.
Recommendation: Use Drägersorb 800 Plus or Drägersorb Free.
3. Fit a new dust filter (C) with every change of soda lime. Only use dust filters from
the list of accessories.
Only use undamaged filters.
4. Attach the CO
2 absorber (D) to the breathing system from below. Rotate it in the
direction of the arrow until it reaches the stop.
Follow the instructions for use for the particular soda lime.
38609
CAUTION
Risk of hypoventilation
Reuse of the dust filter can increase the filter resistance and impair the
ventilation function of the device.
Replace the dust filter with every change of soda lime.
38627
A
B