Reprocessing instructions | Perseus A500 33
Reprocessing
Procedure:
1. Sterilize the components (maximum 134 °C (273.2 °F), 5 min).
2. Check the components for visible damage and replace if necessary.
3.7 After reprocessing
3.7.1 Assembling and fitting device-specific components
Prerequisites:
All components have been reprocessed and dried.
Inspect the following components for damage and wear:
Upper part of the breathing system
Lower part of the breathing system
Incident flow mesh in the inspiratory limb of the lower part of the breathing
system
APL valve
Inspiratory port
Expiratory port
Incident flow mesh in the expiratory port
Inspiratory valve (red)
Expiratory valve (blue)
TurboVent 2 blower module
Flow sensors
Seals and sealing rings
CLIC adapter (option)
If the incident flow meshes are damaged, contact DrägerService.
The following parts must be free from deposits:
Incident flow meshes in the inspiratory limb of the lower part of the breathing
system and in the expiratory port
Valve plate in the inspiratory valve (red)
Valve plate in the expiratory valve (blue)
If necessary, remove deposits on the valve plates with a soft cloth and
reprocess once more.