Reprocessing instructions | Perseus A500 33
Reprocessing
Procedure:
1. Sterilize the components (maximum 134 °C (273.2 °F), 5 min).
2. Check the components for visible damage and replace if necessary.
3.7 After reprocessing
3.7.1 Assembling and fitting device-specific components
Prerequisites:
– All components have been reprocessed and dried.
– Inspect the following components for damage and wear:
– Upper part of the breathing system
– Lower part of the breathing system
– Incident flow mesh in the inspiratory limb of the lower part of the breathing
system
– APL valve
– Inspiratory port
– Expiratory port
– Incident flow mesh in the expiratory port
– Inspiratory valve (red)
– Expiratory valve (blue)
– TurboVent 2 blower module
– Flow sensors
– Seals and sealing rings
– CLIC adapter (option)
If the incident flow meshes are damaged, contact DrägerService.
– The following parts must be free from deposits:
– Incident flow meshes in the inspiratory limb of the lower part of the breathing
system and in the expiratory port
– Valve plate in the inspiratory valve (red)
– Valve plate in the expiratory valve (blue)
If necessary, remove deposits on the valve plates with a soft cloth and
reprocess once more.