Reprocessing instructions | Perseus A500 7
Reprocessing
3.3.2 Classification of device-specific components
The following classification is a recommendation from Dräger.
Non-critical
– Device surfaces
– Trolley and holders
Semi-critical
– Breathing system
– Housing (upper part and lower part)
–Ports
– APL valve
– Inspiratory valve (red) and expiratory valve (blue)
– Absorber container and absorber insert
– TurboVent 2 blower module
– Flexible breathing bag arm (option)
– CLIC adapter (option)
Critical
The device does not contain any components classified as critical.
3.4 Before reprocessing
Observe before disassembly
1. Switch off the device and all devices connected to it.
2. Disconnect all power plugs.
3.4.1 Patient-specific accessories and consumables
The patient-specific accessories and consumables must be removed from the
device and, if necessary, disassembled.
Reusable products:
● If the reusable product has its own instructions for use, perform reprocessing in
accordance with the separate instructions for use. Further information can be
found in the list of accessories.
● If no separate instructions for use are available for the reusable product, perform
reprocessing in accordance with the instructions in this supplement (see
"Reprocessing procedures", page 32).
Disposable products:
● Dispose of the disposable products.