Reprocessing instructions | Perseus A500 7
Reprocessing
3.3.2 Classification of device-specific components
The following classification is a recommendation from Dräger.
Non-critical
Device surfaces
Trolley and holders
Semi-critical
Breathing system
Housing (upper part and lower part)
–Ports
APL valve
Inspiratory valve (red) and expiratory valve (blue)
Absorber container and absorber insert
TurboVent 2 blower module
Flexible breathing bag arm (option)
CLIC adapter (option)
Critical
The device does not contain any components classified as critical.
3.4 Before reprocessing
Observe before disassembly
1. Switch off the device and all devices connected to it.
2. Disconnect all power plugs.
3.4.1 Patient-specific accessories and consumables
The patient-specific accessories and consumables must be removed from the
device and, if necessary, disassembled.
Reusable products:
If the reusable product has its own instructions for use, perform reprocessing in
accordance with the separate instructions for use. Further information can be
found in the list of accessories.
If no separate instructions for use are available for the reusable product, perform
reprocessing in accordance with the instructions in this supplement (see
"Reprocessing procedures", page 32).
Disposable products:
Dispose of the disposable products.

Other manuals for Dräger Perseus A500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Perseus A500 and is the answer not in the manual?

Dräger Perseus A500 Specifications

General IconGeneral
BrandDräger
ModelPerseus A500
CategoryMedical Equipment
LanguageEnglish

Related product manuals