24 Reprocessing instructions | Perseus A500
Reprocessing
3.5.3 Machine cleaning with thermal disinfection
Use a washer-disinfector that meets the requirements of the standard ISO 15883.
Dräger recommends the use of a load carrier for anesthesia accessories and
ventilation accessories. Follow the manufacturer's instructions for the washer-
disinfector.
Components:
– Lower part of the breathing system
– Upper part of the breathing system
– Inspiratory port and expiratory port
–APL valve
– Inspiratory valve (red) and expiratory valve (blue)
– Absorber container and absorber insert
– Flexible breathing bag arm (option)
– TurboVent 2 blower module
– CLIC adapter (option)
Prerequisites:
– The washer-disinfector has been prepared in accordance with the
manufacturer's instructions.
NOTE
Breathing system
At least 250 reprocessing cycles are possible.
NOTE
TurboVent 2 blower module
Number of reprocessing cycles:
– For machine reprocessing, at least 30 reprocessing cycles are possible.
– For manual reprocessing, at least 60 reprocessing cycles are possible.
Step
Components
1)
1) For further information see: "Classification of device-specific components", page 7.
Agent Manufac-
turer
Concentra-
tion
Tempera-
ture
Contact
time
Cleaning All semi-critical
components
neo-
disher
Medi-
Clean
forte
Dr. Weigert Min. 0.30 % Min. 55 °C
(131 °F)
Min. 10 min
Flushing All semi-critical
components
Deminer-
alized
water
–– Tap water
temperature
Min. 1 min
Disinfecting All semi-critical
components
–– – Min. 90°C
(194 °F)
Min. 5 min
Drying All semi-critical
components
– – – – Drying time
depends on
the load