40 Reprocessing instructions | Perseus A500
Reprocessing
3.7.1.3 Inserting the breathing system
1. Insert the assembled breathing system (A) vertically into the breathing system
mount.
2. Turn the lever (B) clockwise by approx. 120°. The breathing system is now
locked.
3. Put the breathing system cover (C) in place and click it into position.
386813868238839