Factory default settings
Dräger Polytron 5100 25
6 Factory default settings
6.1 Settings which can be changed
via the menu
Menu Factory Default Setting
A1 Set depending on the sensor
A2 Set depending on the sensor
A1 RIS / FALL rising (except O2 falling)
A2 RIS / FALL rising
A1 Lat / nLat nLat
A2 Lat / nLat Lat
A1 Ack / nAck Ack
A2 Ack / nAck nAck
A1 ENRG ON / OFF wired in
Docking Station
energized / NO
(normally open)
A2 ENRG ON / OFF wired in
Docking Station
energized / NO
(normally open)
Cal Set steady
Pass Set 0000
LCD Set ON / OFF on
FAIL FREQ PULS / STDY PULS (pulsing)
6.2 Fixed settings
Fault Meaning
Fault Relay
(cannot be
changed) / wired in
Docking Station
energized / NO (normally open)
Yellow Fault LED
Lit if a Fault message is available, see
Section 5.3 on Page 22.
Red Alarm LED
Single blink if A1 condition is present.
Double blink if A2 condition is present.
If an alarm is configured acknowledge-
able and the alarm is acknowledged,
single/double blink changes into
continuous lit.
Alarm Hierarchy
A2 overrides A1 on the LED. However,
the
A1 and A2 relays operate
independently. I.e. if A1 is acknow-
ledgeable, A2 is not, and the gas
concentration is such that it triggers A1
and A2: acknowledging will cause the A1
relay to release. However, the red LED
will still double blink as long as the A2
condition continues to exist.
www.norrscope.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 5100 and is the answer not in the manual?

Dräger Polytron 5100 Specifications

General IconGeneral
BrandDräger
ModelPolytron 5100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals