Charging
the
battery
for
emergency
operation
Primus
has
a
built-in
uninterruptible
power
supply
UPS
which
maintains
the
power
supply
for
at
least
30
minutes
(up
to
90
minutes,
depending
on
the
ventilation
parameters)
in
the
event
of
a
mains
power
failure,
provided
that
the
battery
is
charged.
Switching
to
battery
power
UPS
takes place
automatically
and
is
indicated
on
the
screen
by
the
message:
POWER
FAIL.
The
battery
recharges
automatically
when
the
workstation
is
plugged
into
the
mains,
but
only
up
to
a
maximum
ambient
temperature
of
35
°C.
The
battery
must
be
charged
for
10
hours
before
using
the
workstation
for
the
first
time:
@
Plug
the
mains
plug
of
the
Primus workstation
into
the
mains
socket.
The
mains
voltage
must
correspond
to
that
specified
on
the rating
plate.
The
green
LED
»2-«
lights
up.
@
Leave
Primus
connected
to
the
mains
for
10
hours.
The
workstation
does
not
have
to
be
switched
on.
-
The
devices
connected
to
auxiliary
power
sockets
will
not
be
powered
by
the
UPS
in
the
event
of
a
power
failure.
When
Primus
is
not
in
use
@
The
battery
must
be
charged
at
least
every
4
weeks.
Allowing
it
to
run
low
can
lead
to
damage.
If
Primus
is
not
used
for
an
extended
period:
©
Unplug
the
medical
gas
hoses
from
the wall
supply
points
of
the
central
gas
supply.
@
Close
the
cylinder
valves
on
the
backup
gas
cylinders.
@
Leave
the
workstation
connected
to
the
mains
at
all
times.
1
The
green
LED
»-©«
lights
up.
26
Primus
OOO
DO
On
000000
Ίο
O
Instructions
for
Use
Primus
SW
2.n
uni
0005

Other manuals for Dräger Primus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus and is the answer not in the manual?

Dräger Primus Specifications

General IconGeneral
TypeAnesthesia Workstation
Oxygen FlushYes
Ventilation ModesPressure Support
Gas SupplyO2, Air, N2O
VaporizersIsoflurane, Sevoflurane, Halothane
Patient MonitoringAirway Pressure
Backup BatteryYes
AlarmsApnea, Gas supply failure
Power SupplyAC Mains, Battery Backup

Related product manuals