Ventilation
Ventilating
children
For
tidal
volumes
VT
of
less
than
200
mL:
@
Use
paediatric
hoses.
Connecting
the
breathing
hoses
1
Use
a
Y-piece
with
connection
for
sample
line.
@
The
inspiratory
and
expiratory
microbial
filter
654
St
should
not
be
used
this
reduces
system
compliance.
1
—T
2
Connect
a
0.5
L
breathing
bag
with
socket
to
the
breathing
hose
with
the
large
connection
sleeves.
Slip
the
breathing
hose
over
the
angled
socket.
Hang
the 0.5
L
breathing
bag
onto
the
hook.
3
Slip
the
breathing
hoses
with
the
large
sleeves
onto
the
inspiratory
and
expiratory
sockets
and
connect
the
small
sleeves
to
the
Y-piece.
uni_0039
4
Connect
the
sample
line
to
the
Y-piece
and
water
trap.
uni_0040
To
determine
system
compliance
and
leakage:
5
Firmly
connect
the
Y-piece
to
the
cone.
©
To
determine
system
compliance
and
leakage,
|
|
)
see
"Leak
test"
on
page
73.
uni
0041
69
instructions
for
Use
Primus
SW
2.n