EasyManuals Logo

Dräger Primus User Manual

Dräger Primus
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
Ventilation
Ventilating
children
For
tidal
volumes
VT
of
less
than
200
mL:
@
Use
paediatric
hoses.
Connecting
the
breathing
hoses
1
Use
a
Y-piece
with
connection
for
sample
line.
@
The
inspiratory
and
expiratory
microbial
filter
654
St
should
not
be
used
this
reduces
system
compliance.
1
—T
2
Connect
a
0.5
L
breathing
bag
with
socket
to
the
breathing
hose
with
the
large
connection
sleeves.
Slip
the
breathing
hose
over
the
angled
socket.
Hang
the 0.5
L
breathing
bag
onto
the
hook.
3
Slip
the
breathing
hoses
with
the
large
sleeves
onto
the
inspiratory
and
expiratory
sockets
and
connect
the
small
sleeves
to
the
Y-piece.
uni_0039
4
Connect
the
sample
line
to
the
Y-piece
and
water
trap.
uni_0040
To
determine
system
compliance
and
leakage:
5
Firmly
connect
the
Y-piece
to
the
cone.
©
To
determine
system
compliance
and
leakage,
|
|
)
see
"Leak
test"
on
page
73.
uni
0041
69
instructions
for
Use
Primus
SW
2.n

Other manuals for Dräger Primus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus and is the answer not in the manual?

Dräger Primus Specifications

General IconGeneral
TypeAnesthesia Workstation
Oxygen FlushYes
Ventilation ModesPressure Support
Gas SupplyO2, Air, N2O
VaporizersIsoflurane, Sevoflurane, Halothane
Patient MonitoringAirway Pressure
Backup BatteryYes
AlarmsApnea, Gas supply failure
Power SupplyAC Mains, Battery Backup

Related product manuals