EasyManuals Logo

Dräger Primus User Manual

Dräger Primus
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
Ventilation
Display
(example):
The
airway
pressure
PAW
and
the
Ext.
Outlet
mandatory
frequency
Freg.,
PPEAK
and
PMEAN
are
measured
at
the
external
fresh-gas
outlet.
Pressure
measurement
may
be
impaired
by
activating
the
O2
flush
or
O2
emergency
delivery.
The
minute
volume
MV
and
tidal
volume
VT
are
not
measured.
@
Set
the
fresh-gas
flow.
The
fresh-gas
supply
must
be
equal
to
at
least
twice
the
minute
volume
in
order
to
exclude
rebreathing.
Certain
alarms
are
disabled
automatically
in
order
to
avoid
artefacts,
see
table
on
page
97.
Caution!
Ambient
air
may
become
contaminated
with
anaesthetic
agent
when
using
non-rebreathing
systems.
Excess
fresh gas can
be
discharged
into
the
anaesthetic
gas
scavenging
line
via
the
breathing
system
of
Primus.
For
this
purpose,
the
non-rebreathing
system
must
be
connected
to
the
Y-piece
of
the
breathing
hoses
connected
to
the
breathing
system.
Ending
the
external
fresh-gas
mode
@
Press
any
ventilation
mode
key.
@
The
LED
of
the
selected
ventilation
mode
and
the
display
in
the
status
line
flash.
©
Confirm
via
rotary
knob.
Ventilation
via
the
internal
rebreathing
system
in
Primus
is
restored
directly
in
this
way.
When
changing
from
the
external
non-
rebreathing
system
to
the
rebreathing
system
in
Primus:
@
Reconnect
the
sample
line
to
the
Y-piece.
68
Instructions
for
Use
Primus
SW
2.n
0029

Other manuals for Dräger Primus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus and is the answer not in the manual?

Dräger Primus Specifications

General IconGeneral
TypeAnesthesia Workstation
Oxygen FlushYes
Ventilation ModesPressure Support
Gas SupplyO2, Air, N2O
VaporizersIsoflurane, Sevoflurane, Halothane
Patient MonitoringAirway Pressure
Backup BatteryYes
AlarmsApnea, Gas supply failure
Power SupplyAC Mains, Battery Backup

Related product manuals