Adjustment
ranges
and
default
values
upon
delivery
Ventilation
parameters
Adjustment
range
Default
value
upon
delivery!)
[sec.]
Pressure
limitation
10
to
70
40
PMAX
[mbar]
min.
PEEP
+10
Tidal
volume
VT
[mL]
20
to
14002
600
Frequency
Freq.)
3to
80
12
[1/min]
TINSP”
[sec.]
0.2
to
6.7
1.7
Insp.
pause
time
:
O
to
60
10
insp.
time
TIP
:
TINSP
[%]
PEEP
[mbar]
O
to
20
0
max.
PMAX
-
10
Trigger
sensitivity
OFF,
3.0
(Press.
»Trigger«
[L/min]
0.3
to
15
Supp.
Mode)
OFF
(Vol./
Press.
Mode)
Pressure
support
OFF,
»APPS«®)
[mbar]
3
to
50
5
(Press
Supp.
Mode)
max.
PMAX-PEEP
OFF
(Vol./
Press.
Mode)
Rise
time
«TSLOPE«
0.0
to
2.0 0.0
1)
The
default
values
can
be
set
specifically
for
the
hospital
concerned,
see
page
106.
2)
Optionally
10
to
1400
mL.
3)
Depending
on
the
configuration,
the
inspiratory
time
TinsP
can
be
automatically
changed
together
with
adjustment
of
the
frequency
so
that
the
resultant
ratio
of
inspiration
to
expiration
E
remains
constant.
Only
applies
if
trigger
=
OFF,
see
"Ventilator
and gas
delivery"
on
page
113.
4)
The
resultant
ratio
of
inspiration
to
expiration
|:
E
is
also
displayed
in
parallel.
5)
Optional.
Instructions
for
Use
Primus
SW
2.n
Ventilation
57

Other manuals for Dräger Primus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus and is the answer not in the manual?

Dräger Primus Specifications

General IconGeneral
TypeAnesthesia Workstation
Oxygen FlushYes
Ventilation ModesPressure Support
Gas SupplyO2, Air, N2O
VaporizersIsoflurane, Sevoflurane, Halothane
Patient MonitoringAirway Pressure
Backup BatteryYes
AlarmsApnea, Gas supply failure
Power SupplyAC Mains, Battery Backup

Related product manuals