Ventilation
Adjustment
ranges
and
default
values
upon
delivery
Ventilation
parameters
|
Adjustment
range
Default
value
upon
delivery!)
TSLOPE
[sec.]
Minimum
frequency?)
OFF,
3 to
20
3
FreqMIN
(1/min]
PEEP
[mbar]
0
to
20
0
Trigger
sensivity
0.3
to
15
3.0
Trigger
[L/min]
Pressure
support
8
to
50
5
APPS
[mbar]
Rise
time
0.0
to
2.0
0.0
1)
The
default
values
can
be
set
specifically
for
the
hospital
concerned,
see
page
106.
2)
The
inspiratory
time
is
limited
by
adjustment
of
Freq.MIN
to
yield
a
maximum
ratio
of
1:1
for
|:
E,
thus
ensuring
an
adequate
expiratory
time.
62
Instructions
for
Use
Primus
SW
2.n

Other manuals for Dräger Primus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus and is the answer not in the manual?

Dräger Primus Specifications

General IconGeneral
TypeAnesthesia Workstation
Oxygen FlushYes
Ventilation ModesPressure Support
Gas SupplyO2, Air, N2O
VaporizersIsoflurane, Sevoflurane, Halothane
Patient MonitoringAirway Pressure
Backup BatteryYes
AlarmsApnea, Gas supply failure
Power SupplyAC Mains, Battery Backup