Changing
patients
Changing
patients
To
switch
Primus
to
standby:
@
Press
the
standby
key
Ol
and
confirm
with
the
rotary
knob.
The
functions
of
the
workstation
are
switched
off.
The
set
patient
age,
weight,
alarm
limits,
gas
delivery
settings
and
ventilation
parameters
are
retained.
To
activate
the
default
settings:
©
Press
the
soft
key
»restore
default
settings«
and
confirm.
The
default
settings
for
gas
delivery,
ventilation
parameters
and
alarm
limits
are
restored.
Change
soda
lime
If
the
soda
lime
in
the
absorber
has
turned
violet.
Ifthe
inspiratory
CO2
concentration
is
5
mmHg
or
more.
©
Press
the
standby
key
O)
and
confirm
via
the
rotary
knob.
-A
Slide
the
writing
top
inwards.
2
Press
the
release
button”
on
the
ventilator
unit
and
pull
the
unit
out.
3
Turn
the
absorber
anticlockwise
and
pull
it
down
and
off.
@
Empty
out
the
used
soda
lime
and
dispose
of
it
with
domestic
waste.
@
Fillthe
absorber
to
the
upper
mark
with
fresh
soda
lime.
©
Fitthe
absorber
into
the
breathing
system
from
below
and
turn
it
clockwise
as
far
as
it
will
go.
@
Push
the
breathing
system
inwards
until
it
clicks
into
place.
©
Puli
the
writing
table
out.
not
yet
available
72
Primus
SW
version
20
Last
šelf test:
12
Jun.
00:30
Run
the
self test
at
least
every
24
hours.
The
davice
is
tested
and
ready
for
Enter age and
ideal
body
weight
to
adjust
ventilator
and
alarm
settings.
0020
uni_0042
uni
0043
Instructions
for
Use
Primus
SW
2.n

Other manuals for Dräger Primus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus and is the answer not in the manual?

Dräger Primus Specifications

General IconGeneral
TypeAnesthesia Workstation
Oxygen FlushYes
Ventilation ModesPressure Support
Gas SupplyO2, Air, N2O
VaporizersIsoflurane, Sevoflurane, Halothane
Patient MonitoringAirway Pressure
Backup BatteryYes
AlarmsApnea, Gas supply failure