Changing
patients
End
of
operation
To
switch
Primus
to
standby:
@
Press
the
standby
key
©),
and
confirm
via
the
rotary
knob.
The
workstation
is
now
in
standby.
The
fresh-gas
flow
is
switched
off.
Manual
ventilation
is
possible!
To
switch
Primus
off:
©
Push
the
power
switch
»
Om
«in
completely.
Primus
has
a
power-down
delay.
When
the
power
switch
is
pressed,
an
acoustic
signal
sounds
and
for
10
sec.
the
following
message
is
displayed:
»Please
wait
while
Primus
shuts
down.
Make
sure
that
the
O2
safety
control
valve
is
closed.«
During
this
time,
Primus
can
be
restarted
immediately
by
pressing
the
power
switch
again.
@
Unplug
the
gas
supply
hoses
from
the
wall
sockets.
@
Close
the
cylinder
valves.
If
the
hoses
remain
connected
to
the
wall
sockets,
minute
internal
leaks
may
lead
to
contamination
of
the
gases.
@
Leave
Primus
plugged
into
the
power
supply
in
order
to
charge
the
uninterruptible
power
supply
UPS.
@
We
strongly
recommend
switching
off
the
device
once
a
day
in
order
to
carry
out
the
power-on
self
test.
Remove
any
water
which
may
have
accumulated
in
the
ventilator
diaphragm.
Larger
quantities
.of
condensation
may
impair
operation
of
the
workstation
and/or
lead-to
failure
of
the
eguipment.
Switch
off
the
anaesthetic
gas
scavenging
system
AGS:
©
Disconnect
the
anaesthetic
gas
scavenging
hose.
74
Instructions
for
Use
Primus
SW
2.n
0032