Loading...
Applying
the
Durasensor
DS-100
A
Reusable
sensor
for
short-term
monitoring
ot
relatively
quiet
patients
weighing
over
40
kg.
The
sensor
is
preferably
positioned
on the
index
finger,
although
other
fingers
can
also
be
used.
The
little
finger
should
be
used
if
the
patient
is
particularly
large
or
obese.
@
Open
the
clip
slightly
and
slide
the
sensor
onto
the
finger.
The
tip
of
the
finger
must
touch
the
end
of
the
sensor
and
the
soft
padding
should
rest
on
the
nail
and
tip
of
the finger.
The
lead
should
be
on
top
of
the
finger.
@
Ensure
that
the
finger
is
not
compressed
or
hurt
by
the
clip.
@
Change
the
application
site
after
not
more
than
4
hours
in
order
to
avoid
a
build-up
of
blood
pressure (blocked
circulation).
Follow
the
specific
Instructions
for
Use
when
using
other
Nellcor
sensors!