PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
6 - 9
.
Calibration
Qualified service personnel must completely check the equipment at least annually. Calibration must be
performed in accordance with the Maintenance Interval Table for Resuscitaire products. For details, refer
to the service documentation.
Repair
For repairs, Dräger recommends DrägerService and the use of original Dräger parts.
Maintenance Intervals
When Necessary
Between Patients
Weekly
Every Three Months
Every Six months
Once a Year
Every Two Years
Every Six Years
Personnel Responsible
Replaceable parts:
Mattress X Users and service personnel
Skin probes X X Users and service personnel
Probe covers X X Users and service personnel
Suction bottle X X Users and service personnel
Maintenance:
Device servicing and main-
tenance
X Service personnel
Replace Oxygen hose (wall
or cylinder to Resuscitaire)
X X Users and service personnel
Replace Air hose (wall or
cylinder to Resuscitaire)
X X Users and service personnel
Calibration:
Resuscitation module (if
installed)
(Module must be returned to
vendor for calibration.)
XExpert
Scale (if installed) X X Users and service personnel
Blender (if installed)
(Module must be returned to
vendor for calibration.)
XExpert