PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 25
Alarms
Feature Specification
High Temperature Activates if the skin probe is attached and the skin probe
reaches 39.0°C ±0.2°C (102.2°F). Resets at 38.5°C (101.3°F).
Check Patient Activates in Manual Mode after 10 min. Remains on with
acoustic alarm signal every 30 s for 5 min for a total of 15 min.
The heater then turns off.
APGAR Timer Activates at the 1 min, 5 min, and 10-min APGAR timer inter-
vals.
Power Fail Activates when there is a loss of power. Stays activated for
>10 min.
Probe Fail Activates if the skin probe fails (open or short).
System Fail Indicates a system failure. Refer the unit to service immediately.
Baby Temperature Activates if the baby temperature fluctuates 1 °C above or
below the setpoint.
Electrical Module, Audio Ramping
Alarms (80 dBa maximum for infant)
Tone frequency: 1.2 kHz maximum with a three-stage sound
level: 15 s low, 15-s medium, and then high.
Blender Module, Pneumatic Audio
Alarm Type
Vibrating reed, pneumatically powered.