Dräger Resuscitaire WBR82 - Optional Features (Resuscitaire Birthing Room Warmer (WBR82) and Resuscitaire Radiant; Warmer (RW82) Only)

Dräger Resuscitaire WBR82
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 22
Optional Features (Resuscitaire Birthing Room Warmer (WBR82) and
Resuscitaire Radiant Warmer (RW82) only)
Feature Dimension
RESUSCITATION
Wall Supply Pressure 40 psi (276 kPa) to 75 psi (517 kPa)
Cylinder Pressure 3000 psi (20 684 kPa) maximum
Infant Resuscitator Feature
Patient Gas Supply Airway Pressure Limit,
Operator Adjustable
0cmH
2
O to 50 cmH
2
O (0 kPa to 4.9 kPa) ± 10%
of maximum setting
Patient Outlet Flow Control Range 0 LPM to 15 LPM ± 3% of full scale or ± 10%of
setting, whichever is greater
Fixed Airway Pressure Limit, Preset 60 cmH
2
O (4.9 kPa) ± 20% of setting
Patient Outlet 2 Pressure Limit (MU25175,
MU25179, and MU25180)
160 cmH
2
O (15.7 kPa) ± 10% of setting
Patient Outlet 2 Pressure Limit (MU25175 and
MU25178)
40 cmH
2
O (3.9 kPa) ± 20% of setting
Flow Rate 2 Control Range 0 LPM to 15 LPM ± 3% of full scale or ± 10%of
setting, whichever is greater
AutoBreath™ Infant Resuscitator Feature
(factory installed option)
I:E Ratio Fixed internally at 1:2 nominal (non-adjustable)
Adjustable PEEP at 10 LPM, Minimum PEEP ≤ 0.4 kPa (4 cmΗ2Ο)
/ at 15 LPM, Maximum PEEP > 1.4 kPa
(14 cmH
2
O)
Adjustable Respiratory Rate Range 18 BPM to 60 BPM ± 10% of setting
Adjustable Airway Pressure Relief (range of
working pressure)
0cmH
2
O (0 kPa) to 50 cmH
2
O (4.9 kPa) ±
5cmH
2
O ( 0.5 kPa)
Fixed Maximum Pressure (P Lim max) 50 cmH
2
O ± 10% (5.0 kPa ± 10%)
Fixed Minimum Pressure (P Lim min) 0 kPa
Logic Circuit O2 Consumption 5 LPM
Dead Space 3.4 mL
NOTE: A one-way valve is installed at the Patient Outlet connections. This valve opens when pres-
sure in the hose delivering the gas to the patient falls below -0.4 kPa (-0.06 psi).
Patient Weight Range 10 kg maximum
Gas Supply, O2 Consumption 50 LPM maximum for O2, 20 LPM maximum for
Air
Suction
Suction Circuit Adjustable Suction Intensity 0 kPa (0 mm Hg) to 19.99 kPa (150 mm Hg)
± 0.53 kPa (3.98 mm Hg)
Suction Circuit Maximum Vacuum -19.99 kPa (-150 mm Hg/-2.90 psi) ± 0.53 kPa
(3.98 mm Hg/0.08 psi)
Suction Circuit Maximum Flow Rate < 20 LPM (minimum of 16 LPM at maximum)