PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
2 - 19
Specifications
Standard Features
Feature Dimension
POWER REQUIREMENTS
120 V Models 120 V, 50/60 Hz, 750 W (non-VHA), 1300 W
(VHA)
230 V Models 230 V, 50/60 Hz, 750 W (non-VHA), 1300 W
(VHA)
OVERLOAD PROTECTION
120 V Models 10 A (non-VHA), 10 A (VHA)
230 V Models 10 A (non-VHA), 10 A (VHA)
CHASSIS LEAKAGE CURRENT
120 V Models 300 µA maximum
230 V Models 500 µA maximum
EXAMINATION LIGHT
>100 Foot Candles (0.11 lumens/cm
2
) (50 W
Quartz bulb)
DISPLAYS
Skin Temperature Display Range 18 °C to 43 °C (64.4 °F to 109.4 °F)
Skin Temperature Display Accuracy ± 0.2 °C for 31 °C to 38 °C (88 °F to 100.4 °F)
Skin Temperature Display Resolution 0.01 °C (0.1 °F)
APGAR Timer Display Range 0:00 to 59:59 (min:sec)
APGAR Timer Display Accuracy 0 ± 1 s
APGAR Timer Display Resolution 1 s
MANUAL HEAT CONTROL Adjustable in 10% increments from zero to full
power (100%).
Heater Element Life Expectancy 1000 hours of operation
IRRADIANCE (Pre-Warm Mode)
30% Heater Power Radiant Energy Output ≤10 mW/cm2 maximum
60% Heater Power Radiant Energy Output ≤30 mW/cm2 maximum
100% Heater Power Radiant Energy Output ≤45 mW/cm2 maximum
DATA PORT 2400 bits/s fixed baud rate, RS-232C compatible
MATTRESS TILT (WBR82 only) 0 °, 5 °, and 10 °
BASSINET TILT- Vari-tilt (RW82 only) ± 10 ° from horizontal, ± 1 ° tolerance
DIMENSIONS AND WEIGHT
Mattress Height (from floor) 99.1 cm maximum (39" maximum)
Mattress Height w/VHA (RW82 only) (from
floor)
89 cm minimum to 109 cm maximum (35" mini-
mum to 43" maximum)
Mattress Length 66.0 cm (26”)
Mattress Width 53.3 cm (21”)
Mattress Thickness 2.54 cm (1”)
Height (WBR82) 188 cm (74")