PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
4 - 19
5. Check the pass-through drawer (Resuscitaire Birthing
Room Warmer shown):
a. Slide the pass-through drawer in and out on both
sides of the cart and bassinet.
b. Return the pass-through drawer to its center
position.
6. Check the warmer module swivel operation:
WARNING:
Push the sides of the warmer module. Do not touch
under the warmer module. Personal injury could
occur.
a. Push the sides of the warmer module, and rotate it
90 ° to the left or right of center. Do not touch
under the warmer module.
WARNING:
For effective heating, ensure that the warmer module
is centered above the patient. Failure to do so could
result in personal injury.
b. Return the warmer module to its center position.
7. Check the operation of the accessory x-ray tray:
a. Grasp the middle of a side panel, and pull the x-ray
tray out from under the bassinet.
b. Return the x-ray tray to under the bassinet.