Menu functions
Dräger X-am 5100 13
4 Menu functions
4.1 Calling the Info Mode
In measuring mode, press the key for approx. 3 seconds.
If any warning or fault messages are present, the
corresponding information or error codes are displayed
(see "Troubleshooting" on page 19 to page 20).
Press the key successively for the next display.
The peak values and the exposure values TWA
1
and
STEL
1
are displayed.
If no key is pressed for 10 seconds, the instrument reverts
automatically to measuring mode.
4.2 Calling the Info-Off Mode
When the instrument is in a deactivated state, press the
key.
The name of the gas, measuring unit and measuring
range limit value are displayed.
Pressing the key again exits the Info Off mode (or via
timeout).
4.3 Quick Menu
4.3.1 Quick menu functions
4.3.2 Calling the Quick Menu
The fresh air adjustment and the bump test are activated in the
quick menu on delivery. The function for displaying and
deleting peak values can additionally be activated with the
Dräger CC Vision PC software.
1. In measuring mode, press the key three times.
If functions in the quick menu have been activated using
the Dräger CC-Vision PC software, you can select these
functions using the key. If no functions have been
activated in the quick menu, the instrument remains in
measuring mode.
2. You can select the activated functions of the quick menu by
pressing the key.
Press the key to call the selected function.
Press the key to cancel the active function and to
switch to measuring mode.
If no key is pressed for 60 seconds, the instrument
reverts automatically to measuring mode.
4.3.3 Quick menu "Displaying and deleting peak values"
After the function has been
selected, the current peak
values are displayed;
the special peak values
symbol appears in the
display at the same time.
1. The peak values can be
deleted by pressing the
key for 5 seconds.
The next display appears.
2. Press the key to end
the function.
1 Only when activated in the instrument configuration. Delivery status:
not activated.
Warning messages are displayed. Numerical codes
of warning messages: see "Warning messages" on
page 19.
key
Fault messages are displayed. Numerical codes of
fault messages: see "Fault messages" on page 20.
key
The peak values = the maximum measured values
are displayed.
key
The average values of the exposures based on
a shift of, e.g., 8 hours (TWA) are displayed
key
The short-term values (STEL) = average values of
the concentrations over the average value duration
are displayed
key
The instrument is in measuring mode again
OK
OK
OK
OK
OK
OK
OK
OK
Bump test, see "Performing the bump test" on page 11.
Fresh air adjustment, see "Performing a fresh air
adjustment" on page 15.
Display and deletion of the peak values, see "Quick
menu "Displaying and deleting peak values"" on
page 13.
OK
10133280.eps
HCl ppm
12
10233280.eps
HCl ppm
0
OK
OK