20 Dräger X-am 5100
Troubleshooting
7.2 Fault messages
Special symbol » « and
displayed numerical code:
Cause Remedy
102 The customer's service life counter
has elapsed
Reset the service life counter using Dräger
CC Vision.
103 The instrument is defective The instrument must be repaired by
DrägerService.
104 Check sum error program code The instrument must be repaired by
DrägerService.
105 The bump test interval has elapsed Carry out the bump test, see "Calibrating the
instrument" on page 15.
106 The calibration interval has elapsed
(at least 1 calibration interval has elapsed)
Perform a sensitivity adjustment,
see "Calibrating the instrument" on page 15.
107 Bump test error Perform bump test, see "Performing the
bump test" on page 11, or snsitivity
adjustment, see "Calibrating the instrument"
on page 15.
108 The instrument is defective The instrument must be repaired by
DrägerService.
109 The menu function cannot be carried out
because of an error.
Identify the error code via the info menu
and switch off, if necessary.
301 No valid zero point adjustment of the
Dräger Sensor XS EC1
Perform a fresh air calibration,
see "Performing a fresh air adjustment" on
page 15.
302 No valid sensitivity adjustment of the
Dräger Sensor XS EC1
Perform sensitivity adjustment or fresh air
adjustment, see "Calibrating the instrument"
on page 15.
303 The measured value of DrägerSensor XS
EC 1 is in the negative range
Perform a fresh air calibration,
see "Performing a fresh air adjustment" on
page 15.
304 Dräger Sensor XS EC1 not inserted or
defective
Check DrägerSensor XS EC1,
see "Replacing the Sensors" on page 18.
305 Error during bump test of Dräger Sensor
XS EC1
Repeat bump test. If necessary, adjust or
replace the Dräger Sensor XS EC1,
see "Replacing the Sensors" on page 18.
326 Error during accelerated warm-up of
Dräger Sensor XS EC1
Remove the power pack and insert it again or
replace the sensor. Sensor must not be
exposed to gas during the first 5 minutes.