Troubleshooting
Dräger X-am 5100 19
7 Troubleshooting
To display the numerical codes of the warning and fault messages in the info mode, see "Calling the Info Mode" on page 13.
7.1 Warning messages
Fault Cause Remedy
Not possible to switch on
the instrument
Discharge the power pack Charge the power pack, see "Charging the
rechargeable batteries" on page 7.
Discharge the alkaline batteries Insert new alkaline batteries, see "Replacing
the batteries / rechargeable batteries" on
page 8.
Not possible to switch off
the instrument
The instrument is not set to measuring
mode
Select measuring mode.
The instrument is configured to "Disable
prohibited"
Configure the instrument to "Disable allowed"
with Dräger CC Vision.
Display » – – « Measuring range calibrated incorrectly Recalibrate the measuring range,
see "Calibrating the instrument" on page 15.
Electronics or sensors defective Must be repaired by DrägerService.
Special symbol » « and
displayed numerical code:
Cause Remedy
152
Customer's service life counter about
to elapse
Reset the service life counter using
Dräger CC Vision.
153
Database 90 % full Read the database soon and then clear
the memory.
154
Database full Read the database and clear memory.
155
Interval for the bump test has elapsed Carry out the bump test, see "Performing the
bump test" on page 11.
159
Adjustment not possible. The menu
function cannot be carried out because
of a message which is preventing the
function (e.g., sensors in warm-up phase).
Identify the message code via the info menu
and switch off, if necessary.
351
DrägerSensor XS EC1 in the warm-up
phase
Wait until warm-up time is complete.
352
DrägerSensor XS EC1 in the warm-up
phase
Wait until warm-up time is complete.
353
EC1 concentration has drifted into the
negative range
Perform a fresh air calibration,
see "Performing a fresh air adjustment" on
page 15.
354
The temperature is too high Operate the instrument within the allowed
temperature range.
355
The temperature is too low Operate the instrument within the allowed
temperature range.
356
The calibration interval for DrägerSensor
XS EC1 has elapsed
Perform a sensitivity adjustment for
DrägerSensor XS EC1, see "Calibrating the
instrument" on page 15.