18 Dräger X-am 5100
Replacing the Sensors
6 Replacing the Sensors
1. Connect the instrument to the PC with the USB-DIRA
adapter (Order No. 83 17 409).
2. Deactivate the slot using the CC-Vision PC software.
3. 4 screws on the lower shell should now be unfastened.
4. Carefully release the upper shell upwards from the
lower shell. Do not tilt the upper shell when doing this.
5. Follow the remaining instructions in the CC Vision PC
software.
NOTICE
When the upper shell is being released, it is possible
that the DrägerSensor will be pulled out of its mount at
the same time and remain in the upper shell.
00433280.eps
IR
USB 1.1
0
X-am 5100
00133280.eps
i
i