Calibrating the instrument
Dräger X-am 5100 15
5 Calibrating the instrument
Adjustment may not be possible due to instrument and
channel errors.
Allow the sensor to warm up before the adjustment.
Warm-up time: see Instructions for Use/data sheet for the
installed DrägerSensor.
5.1 Adjustment interval:
Observe the relevant specifications in the Instructions for
Use/data sheet for the DrägerSensor installed.
For critical applications according to EN 60079-29-2
1
or
EN 45544-4
2
and national regulations.
Improvement of zero point accuracy – perform a fresh air
adjustment, see "Performing a fresh air adjustment" on
page 15.
5.2 Performing a fresh air adjustment
A fresh air adjustment can be performed to improve the zero
point accuracy.
Calibrate the instrument to fresh air, free of measured
gases or other interfering gases.
The zero point of the sensor is set to 0 during the fresh
air calibration.
Sensors that are faulty or not warmed-up will prevent the
adjustment.
In the case of sensors which are in the warm-up phase,
the message » 159 « is displayed with the special
symbol » « (for warning message).
In the case of a sensor or instrument error, the
message » 109 « is displayed with the special symbol
» « (for a fault message).
The message is cleared after 5 seconds and the
function is available again in the menu.
1. Switch on the instrument.
2. Depending on instrument configuration:
Call up the Quick Menu and select the fresh air
adjustment function » «, see "Quick Menu" on
page 13.
or
Call up the Adjustment menu and select the fresh air
adjustment function » «, see "Adjustment menu" on
page 14.
The measured values
flash.
When the measured
values have stabilized:
3. Press the key to
carry out the fresh air
adjustment.
The display containing the
current gas concentration
changes with the display
» OK «.
4. Press the key to exit
the calibration or wait for
approx. 5 seconds.
If a fault has occurred
during the fresh air adjustment:
The fault message » «
appears and » « is
displayed for the respective
sensor instead of the
measured value.
In this case, repeat the
fresh air adjustment.
If necessary, replace the
sensor, see "Replacing
the Sensors" on page 18.
WARNING
Always adjust the zero point first, before the sensitivity.
Otherwise the adjustment will be incorrect!
1 EN 60079-29-2 – Guidelines for selection, installation, use and maintenance
of instruments for the detection and measurement of flammable gases and
oxygen.
2 EN 45544-4 – Electrical instruments for the direct detection and direct
concentration measurement of toxic gases and vapors – Part 4:
Guidelines for selection, installation, use and maintenance.
!
10733280.eps
HCl
ppm
0.0
OK
10833280.eps
HCl ppm
OK
OK
10933280.eps
HCl ppm
--