14 Dräger X-am 5100
Menu functions
4.4 Adjustment menu
4.4.1 Functions of the adjustment menu
4.4.2 Calling the adjustment menu
The adjustment menu can only be accessed by entering
a password.
Password on delivery: » 001 «
The default password on delivery can be changed using
the PC software Dräger CCVision.
1. In measuring mode, press the key for at least 5 seconds.
The function for entering the password is selected.
The special symbol » « (for the "Enter password"
function) is displayed.
The display shows » 000 «, with the first digit flashing.
2. Use the key to set the
flashing digit.
3. Press the key, the
second digit starts flashing.
4. Use the key to set the
flashing digit.
5. Press the key, the third
digit starts flashing.
6. Use the key to set the
flashing digit.
7. Press the key to
confirm the password once
it has been set completely.
8. The adjustment menu
functions can now be selected by pressing the key.
Press the key to call the selected function.
Press the key to cancel the active function.
If no key is pressed for 10 minutes, the instrument
reverts automatically to measuring mode.
Fresh air adjustment, see "Performing a fresh air
adjustment" on page 15
1-button sensitivity adjustment
Standard sensitivity adjustment
10333280.eps
000
OK
OK
OK
OK