10 Dräger X-am 5100
Use
3.2 Configuration
3.2.1 Standard gas configuration
3.2.2 Standard instrument configuration
1
3.2.3 Configuring the instrument
To customise its configuration, the instrument must be
connected to a PC with the USB-DIRA adapter (Order
No. 83 17 409). Dräger CC-Vision PC software is used to
perform the configuration.
Changing the standard configuration
The installed Dräger CC Vision PC software is used for
configuration.
Observe the documentation and online help of the software.
A version of the CC-Vision software that can be used for
Dräger X-am 5100 is available for download from the
product page for the X-am 5100 at the following web
address: www.draeger.com.
Reading the database and displaying it Graphically
The installed PC software Dräger GasVision is used for
reading and displaying the database.
Observe the documentation and online help of the software.
Changing the configuration: See “Troubleshooting” on
page 19.
DrägerSensor Measuring range
1
Alarm A1
1
Alarm A2
1
threshold
can be acknowledged
self-latching
threshold
can be acknowledged
self-latching
XS EC HF/HCI
[ppm]
2
0 to 30 5 Yes No 10 No Yes
XS EC H
2
O
2
[ppm]
0 to 20 1 Yes No 2 No Yes
XS EC N
2
H
4
[ppm]
0 to 3 0.1 Yes No 0.2 No Yes
1 Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision
software. A version of the CC-Vision software that can be used for Dräger X-am 5100 is available for download from the product page for the X-am 5100 at the
following web address: www.draeger.com
2 Valid for HCl.
1 Different settings can be selected to meet customer requirements on
delivery. The current setting can be checked and changed with the
Dräger CC Vision software.
Dräger X-am 5100
Bump test mode Quick bump test
Fresh air adjustment On
Life signs (visual only) On
Disabling permitted / blocked for A2
Averaging time 15 minutes for STEL
8 hours for TWA
WARNING
After a basic initialization has been carried out with
the Dräger CC Vision PC software, individual alarm
settings may have been changed.
00433280.eps
IR
USB 1.1
0
X-am 5100
!