ObservAir Operating Manual
Flow control
None
Table 5. Serial port settings
3.8.3. Sensor configuration: Serial commands
Commands can be sent over the Serial Monitor to configure the
ObservAir’s various operational settings. Most serial commands share the
same basic format: $cmd$val where ‘cmd’ is the name of command, and
‘val’ is the desired setting value. For example, the command $setFR$100
sets the sample flowrate to 100ccm. Table 6 documents all available
commands. All settings are saved to the SD card for each mission.
Command (cmd)
Value (val)
Function
$setMissonName$val
String
Set mission name (folder name)
$setID$
0 to 1000
Set sensor ID
$setFR$val
30 to 200
Set sample flow rate in ccm
$setSamplePeriod$val
2 to 60
Set measurement period in seconds
$setTimeZone$val
-12 to 12
Set time zone in hours from GMT
$syncTime$
N/A
Sync time with internet server (Time
zone must be set, internet must be
accessible and credentials
provisioned)
$setTime$YYYY,MM,
DD,HH,MM,SS
date/time
Manually set clock to specified
date/time
$deviceInfo$
N/A
Returns all sensor settings
$setMaxATN$val
10 to 150
Set the maximum attenuation (ATN)
value, above which alarm is triggered
$ATNreset$
Reset the attenuation (ATN) to 0
$setLEDdispMode$val
BC, GAS1,
GAS2,
ALERT, OFF
Set the LED to pollutant concentration
display, alerts only, or turn off
$setLEDmaxConc$val
0.2 to 50
Set maximum concentration displayed
by LED button (red)
$setLEDra$val
60 to 3600
Set running average size (in seconds)
for the LED concentration display
$setLEDtimeout$val
0 to 60
LED display timeout duration in
minutes. 0 = LED on indefinitely.
$setDLPF$ val
0 / 1
Toggle the DLP filter applied to the
pollutant concentration data