EasyManua.ls Logo

DST ObservAir - Calibration: Gaseous Pollutants

Default Icon
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
ObservAir
®
Operating Manual
should be checked for leaks (see Section 5.2). If the problem persists after
a successful leak check, contact DSTech for assistance. The ObservAir
should provide a robust zero reading during calibration. Zero-calibration
is straightforward and can be carried out regularly to verify the sensor’s
proper operation.
In order to span-calibrate the BC data, the ObservAir must be
collocated with a reference instrument. If calibrating relative to another
absorption photometer, ensure that the reference instrument is operating
at a wavelength of 880 nm, like the ObservAir. Derive a linear regression
between the ObservAir and reference BC data, as shown in Equation 6,
and note the slope. Generally, this calibration should be done using hourly-
average data collected over at least 1 week of collocated sampling.


󰇛
󰇜


󰇛
󰇜
󰇛󰇜
BC
ref
(t) = Reference BC measurement at time ‘t’ (µg/m
3
)
BC
OA
(t)
= ObservAir BC measurement at time ‘t’ (µg/m
3
)
m = Linear regression slope
b = Linear regression intercept (µg/m
3
)
Using the slope value, adjust the ObservAir’s MAC factor to reconcile the
two datasets, as shown in Equation 7. For example, if the ObservAir’s BC
measurements are consistently 10% lower than the reference (slope of
1.1), the MAC should be decreased by 10% to calibrate the ObservAir data
relative to the reference.




󰇛󰇜
MAC
cal
= Calibrated MAC (m
2
/g)
MAC
default
= Default MAC (m
2
/g) = 7.8 m
2
/g from factory
Note: BC measurements are also proportional to flow rate, so it is crucial
that the ObservAir’s flow rate sensor be calibrated prior to any BC span
calibrations (see Section 4.3).
4.2. Calibration: Gaseous pollutants
The most accurate and practical method for gas sensor calibration is
collocation with a reference instrument in the intended deployment
setting. As discussed in Section 1.2.2, the electrochemical cells
operation varies significantly depending on environmental conditions, so

Table of Contents