EasyManua.ls Logo

Durr Dental Vector - Page 13

Durr Dental Vector
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ambient conditions during transport and storage
Temperature °C -15 to +60
Relative humidity % Max. 95
Air pressure kPa 75 - 106
Electromagnetic compatibility (EMC)
Interference emission measurements
Interference voltage at the power supply connection
CISPR 11:2009+A1:2010
Compliant
Electromagnetic interference radiation
CISPR 11:2009+A1:2010
Compliant
Emission of harmonics
IEC 61000-3-2:2005+A1:2008+A2:2009
Compliant
Voltage changes, voltage fluctuations and flicker emis-
sions
IEC 61000-3-3:2013
Compliant
Electromagnetic compatibility (EMC)
Interference immunity measurements on the cover
Immunity to electrostatic discharge
IEC 61000-4-2:2008
8 CD, 2 kV AD, 4 kV AD, 8 kV AD, 15 kV AD, evaluation
criterion: B
Compliant
Immunity to high-frequency electromagnetic fields
IEC 61000-4-3:2006+A1:2007+A2:2010
80 MHz... 2.7 GHz
3 V/m
80% AM, 1 kHz sine, evaluation criterion: A
Compliant
Immunity to near fields of wireless HF communication
devices
IEC 61000-4-3:2006+A1:2007+A2:2010
Refer to the table with immunity to interference levels for
near fields of wireless HF communication devices.
Compliant
Immunity to power frequency magnetic fields
IEC 61000-4-8:2009
50 Hz, 30 A/m, in x-y-z-direction
Compliant
Electromagnetic compatibility (EMC)
Interference immunity measurements on the supply input
Immunity to fast electrical transients/bursts – AC mains
voltage
IEC 61000-4-4:2012
± 2 kV
100 kHz repetition rate
Compliant
Immunity to interference, surges
IEC 61000-4-5:2005
± 0.5 kV, ± 1 kV, ± 2 kV
Compliant
Product description
9000-615-32/02 2012V003 11
EN

Related product manuals