Electromagnetic compatibility (EMC)
Interference immunity measurements on the supply input
Immunity to conducted disturbances, induced by radio-
frequency fields – AC mains voltage
IEC 61000-4-6:2013
3 V
0.15–80 MHz
6 V
ISM frequency bands
0.15–80 MHz
80% AM at 1 kHz
Compliant
Immunity to voltage dips, short interruptions and voltage
variations
IEC 61000-4-11:2004
Compliant
Immunity to interference table, near fields of wireless HF communication devices
Radio service Frequency band
MHz
Test level
V/m
TETRA 400 380 - 390 27
GMRS 460
FRS 460
430 - 470 28
LTE band 13, 17 704 - 787 9
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE band 5
800 - 960
28
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE band 1, 3, 4, 25
UMTS
1700 - 1990
28
Bluetooth
WLAN 802.11 b/g/n
RFID 2450
LTE band 7
2400 - 2570 28
WLAN 802.11 a/n 5100 - 5800 9