10 Operation
10.1 Display/handling
Scaler handpiece inserted:
POWER and LIQUID adjustment possible
No handpiece inserted:
The settings cannot be changed.
The LEDs are only active and light up
when a handpiece is inserted.
The last settings set up for the handpiece
that is inserted remain active after the unit
is switched off and back on again.
Any running cleaning or rinsing processes
can be stopped by touching the corre-
sponding button again.
LED off
LED on
LED flashing
11 ON / Standby
For switching on or off; touch the
button for at least 2 seconds.
Standby, unit switched off.
ON, unit switched on.
If no function is used for a period of
30 minutes, the unit will automati-
cally switch off (standby).
12
POWER (power setting)
1 – 5 LEDs light up depending on
the power setting selected (5 LEDs
= maximum power)
Tip: The power setting can also be
adjusted during treatment.
13
RINSE (rinsing/disinfecting)
To start the rinsing process: touch
the button for at least 2 seconds.
The blue LED flashes during the
rinsing process.
14 CLEAN
After approx. 30 operating hours the
blue LED will light up continuously –
this shows that it is time for clean-
ing.
To start the cleaning process: touch
the button for at least 2 seconds.
The blue LED flashes during the
cleaning process.
Recommendation: Clean the unit
every four weeks or, at the latest, as
soon as the LED lights up continu-
ously.
15
LIQUID
This display is only active when the
Scaler handpiece is inserted.
1 LED on = minimum fluid con-
sumption (30 ml/minute)
3 LEDs on = maximum fluid con-
sumption (45 ml/minute)
16
PEDAL (flexible foot switch)
LED on: low battery power – change
the battery of the flexible foot
switch.
LED flashing: no flexible foot switch
connected (cable operation) or
assigned (wireless operation).
17 FUNCTION
LED lights up: The operation has
been interrupted. Clean the instru-
ment chuck with the air and water
syringe and dry it, then continue the
treatment.