Proceed as follows:
-Connect ecom-J2KN
pro
and PC via USB cable.
-Select “Load data“ and confirm with <OK>.
-Answer the displayed question with YES (<F1>).
-Decide if the data recorded can be cancelled
(<F1> for YES / <F4> for NO).
-Start the data transfer on your PC.
Send data:
With this function the data records completed with measurement
values can be transferred to the PC programme (procedure similar to
chapter „Load data“).
Format:
This function is usually needed by the initial adjustment of the in-
strument at our factory (preparation of internal memory for data rec-
ord).
Caution: All stored values will be cancelled!
12.2. Automatic measurement (option)
The configuration of the ecom-J2KN
pro
with operation securing, self
regulating additional components like the condense trap, the gas
cooler and the magnetic valve technique permits even a long term
operation of the instrument. By setting time intervals for the meas-
urement the instrument switches autonomously to fresh air purge
after the measurement phase (gas will be sucked via the connection
fresh air) and operates a calibration phase for the sensors.
This cycle repeats itself until the auto-
matic measurement will be finished. For
setting the time intervals select from the
menu “Data processing” the menu
item “Automatic measu.”. The time
intervals have following meanings: