Chapter 13 Monitoring TEMP
13.1 Overview
Body temperature is measured by means of a thermistor probe (a semiconductor whose resistance
changes with temperature) that is applied to the skin or to the rectum.
Two TEMP probes can be used simultaneously to measure two TEMP values, and get the
temperature difference. The standard configuration is skin probe for adult.
13.2 TEMP Safety Information
WARNING
1 Verify probe cables fault detection before the beginning of monitoring phase. Unplug
the temperature probe cable of the channe1 from the socket, and then the screen will
display the error message TEMP T1 Sensor Off and the audible alarm is activated. It
is the same to the other channel.
2 Take the TEMP probe and cable carefully. When they are not in use, you should coil
up the probe and cable into a loose circle. If the wire inside the cable is tensely pulled,
it may cause mechanical damage to the probe and the cable.
3 Temperature probes do not need any probe covers; please remember to disinfect the
probe after each use on a patient.
NOTE:
The reference body site temperature is the same as the temperature of the measuring
site.
Connector for TEMP probe 1
Connector for TEMP probe 2