Patient Monitor User Manual V-Link Module
- 190 -
Click on the Anes/Vent Interface, open the setup menu, the user can:
• Select the parameter to be displayed on the window in the Parameter list.
• Select a suitable ruler for the waveform from the options TopRuler, MidRuler and
BotRuler. MidRuler is not available when anesthesia device is connected.
• Select an appropriate sweep for the waveform in the Sweep list. The bigger the value is, the
wider the waveform is.
• Choose Mode and set it to Curve or Filled from the pop-up list.
• Select Color Setup to make color changes on parameter and waveform.
• Select Default to restore factory default for all setups in this menu.
• Select Enter Full-param Interface to display all the sub-parameters’ measurement values,
without waveforms displayed.
• Select Respiratory Loop to enter respiratory loop interface (only applicable to Maquet
ventilator).
• View the Unit and Baro Pess.
NOTE:
1 When the ventilator is working, it must be ventilated. If not ventilated (air source is
disconnected), there may be a risk.
2 The monitor does not provide alarms from the ventilator itself.
3 When switching between different devices, the trend data of the previous device is
cleared.
21.6.2.1 Respiratory Loop Interface
Normally, click on the Anes/Vent window to open the setup menu, select Loop in the Parameter
list, and user can enter respiratory loop interface by clicking
Large Loop.
Respiratory loops can indicate a fault in the airway tubing and help physicians to detect
respiratory problems of patients. The two types of loops are available in real time:
• F-V (flow-volume) loops: it illustrates the dynamic relation between flow and volume during
respiration and provides information about condition of the airway tubing.
• P-V (pressure-volume) loops: it reflects the dynamic relation between pressure and volume
as well as compliance of the respiratory system.