Patient Monitor User Manual Product Specifications
- 249 -
Pace Pulse
Pulse Indicator
Pulse is marked if the requirements of IEC 60601-2-27: 2011, Sect.
201.12.1.101.12 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to2.0 ms
Ascending time: 10 μs to 100 μs
Pulse Rejection
Pulse is rejected if the requirements of IEC 60601-2-27: 2011, Sect.
201.12.1.101.13 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 μs to 100 μs
Pace Pulse Detecting Lead: one among I, II, III, AVR, AVL, AVF, V1, V2, V3,V4, V5, V6
Heart Rate
HR Calculation
Range
ADU: 15 bpm to 300 bpm
PED/NEO: 15 bpm to 350 bpm
Accuracy
±1% or 1 bpm, whichever is greater
Resolution 1 bpm
Sensitivity 300 μVPP
QRS Detection
Range
The detection range has exceeded the requirement described in the
standard:
Width: 70 ms~120 ms for adult, 40 ms~120 ms for Pediatric/neonate.
Amplitude: 0.5 mv~5 mv
In adult mode, these two signals are not responded:
1. when QRS amplitude of 0.15 mV or less is applied;
2. when QRS duration of 10 ms and QRS amplitude of 1 mV or less is
applied.
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.15.
PVC
Range ADU: (0 to 300) PVCs/ min
PED/NEO: (0 to 350) PVCs/ min
Resolution 1 PVCs/min
Pauses/min
Range ADU/PED/NEO: (0 to 30) pauses/min
Resolution 1 pause/min