Patient Monitor User Manual Product Specifications
- 263 -
Balance Gas Compensation Room air, N
2
O, helium
Stability
Short Term Drift Drift over 4 hours < 0.8 mmHg
Long Term Drift 120 hours
Zero Calibration Support
☆Alarm Type
EtCO
2
, FiCO
2
, AwRR
☆Apnea Alarm Delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Data Sample Rate 100 Hz
CO
2
Rise Time/Response Time
(mainstream)
Less than 60 ms
Sensor Response Time
(sidestream)
< 3 seconds, including transport time and rise time
Interfering Gas and Vapor Effects on EtCO
2
Measurement Values:
Gas or Vapor Gas Level (%) Quantitative Effect/Comments
Nitrous oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Desflurane
60
4
5
5
5
80
50
15
Dry and Saturated Gas
(0 ~ 40) mmHg: ±1 mmHg additional error
(41 ~ 70) mmHg: ±2.5% additional error
(71 ~ 100) mmHg: ±4% additional error
(101 ~ 150) mmHg: ±5% additional error
*Additional worst case error when compensation
for P
B
, O
2
, N
2
O, anesthetic agents, or helium is
correctly selected for the actual fractional gas
constituents present.
Desflurane:
The presence of desflurane in the exhaled breath at
concentrations greater than 5% will positively bias
Carbon Dioxide values by up to an additional 3
mmHg at 38 mmHg.
Xenon:
The presence of Xenon in the exhaled breath will
negatively bias Carbon Dioxide values by up to an
additional 5 mmHg at 38 mmHg.