Patient Monitor User Manual Product Specifications
- 267 -
Accuracy-
All
Conditions
Gas Accuracy
CO
2
± (0.3 kPa + 4% of reading)
N
2
O ± (2 kPa + 5% of reading)
Agents ± (0.2 kPa + 10% of reading)
O
2
± (2 kPa + 2% of reading)
AwRR Accuracy
±1 rpm
Apnea Alarm
Delay
20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Alarm
Providing alarms of EtCO
2
, FiCO
2
, EtO
2
, FiO
2
, EtN
2
O , FiN
2
O ,
EtAA , FiAA , AwRR
Exhaust Emission Interface for exhaust collection is available
Support:
Zero calibration
O
2
compensation
N
2
O compensation
Interfering Gas and Vapor Effects:
Gas or Vapour Gas Level CO
2
Agents N
2
O
ISA AX+
ISA OR+
N
2
O
4
)
60 vol%
_
1
)
_
1
)
_
1
)
HAL
4
)
4 vol%
_
1
)
_
1
)
_
1
)
ENF, ISO, SEV
4
)
5 vol%
_
1
)
_
1
)
_
1
)
DES
4
)
15 vol%
_
1
)
_
1
)
_
1
)
Xe(Xenon)
4
80 vol% -10% of reading
3
_
1
)
_
1
)
He(Helium)
4
50 vol% -6% of reading
3
_
1
)
_
1
)
Metered Dose Inhaler
Propellants
4)
Not for use with metered dose inhaler propellants
C
2
H
5
OH(Ethanol)
4
)
0.3 vol%
_
1
)
_
1
)
_
1
)
C
3
H
7
OH(Isopropanol)
4
)
0.5 vol%
_
1
)
_
1
)
_
1
)
CH
3
COCH
3
(Acetone)
4
1 vol%
_
1
)
_
1
)
_
1
)
CH
4
(Methane)
4
3 vol%
_
1
)
_
1
)
_
1
)
CO(Carbon monoxide)
5
1 vol%
_
1
)
_
1
)
_
1
)