Patient Monitor User Manual Product Specifications
- 272 -
☆AwRR Accuracy
±1 rpm
☆Apnea Alarm Delay
20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Work Mode Measure
☆Alarm
Providing alarms of EtCO
2
, FiCO
2
, EtN
2
O , FiN
2
O , EtAA, FiAA,
AwRR
Support:
Real-time gas concentration monitoring
Zero calibration
Interfering Gas and Vapour Effects:
Gas or Vapour Gas Level CO
2
Agents N
2
O
IRMA AX+
N
2
O
4
60 vol%
1&2
1
1
HAL
4
4 vol%
1
1
1
ENF, ISO, SEV
4
5 vol%
1
1
1
DES
4
15 vol%
1
1
1
Xe(Xenon)
4
80 vol% -10% of reading
3)
1
1
He(Helium)
4
50 vol% -6% of reading
3)
1
1
Metered Dose Inhaler
Propellants
4)
Not for use with metered dose inhaler propellants
C
2
H
5
OH(Ethanol)
4
0.3 vol%
1
1
1
C
3
H
7
OH
(Isopropanol)
4)
0.5 vol%
1
1
1
CH
3
COCH
3
(Acetone)
4)
1 vol%
1
1
1
CH
4
(Methane)
4
3 vol%
1
1
1
CO(Carbon monoxide)
5)
1 vol%
1
1
1
NO(Nitrogen
monoxide)
5)
0.02 vol%
1
1
1
O
2
5
100 vol%
1&2
1
1
Note 1: Negligible interference, effect included in the specification “Accuracy, all conditions”
above.