Patient Monitor User Manual Alarm Information
- 55 -
Message Cause Alarm Level Action Taken
ECG Signal
Exceeded
ECG measuring signal is
beyond measuring range.
Low
Check lead
connection and
patient condition
ECG Noise
ECG measuring signal is
greatly interrupted.
Low
ECG Comm Fail
ECG module failure or
communication failure
High
Stop measuring
function of ECG
module, and notify
biomedical engineer
or manufacturer’s
service staff.
RESP
RESP Comm Fail
RESP module failure or
communication failure
High
Stop measuring
function of RESP
module, and notify
biomedical engineer
or the
manufacturer’s
service staff.
RESP Noise
RR cannot be measured due
to patient movement.
Low
Check whether the
RESP leads are well
connected. Keep the
patient calm for
better monitoring.
RR Exceed
RR measuring value is out
of the measure range.
Medium
Check whether
interference to the
respiratory signal
exists. And check
whether the patient
is breathing
normally; breathing
too rapidly or too
slowly may
endanger patient’s
life.