Patient Monitor User Manual Alarm Information
- 62 -
Message Cause Alarm Level Action Taken
AG AA Id
Unreliable
1) Mainstream: The airway
adapter was replaced
without a zeroing.
2) More than 2 anesthetic
agents are present in the
breathing circuit.
3) High concentrations of
solvents, cleaning agents or
other interfering gases are
present in the breathing
circuit.
Medium
1) Perform a
zeroing after
replacing the
adapter.
2) Reduce the
number of
anesthetic agent
types.
3) Replace the
sampling tube or
reduce the
interfering gases.
AG Replace Adapter
Replacement of the adapter
is required.
Medium Replace the adapter.
AG TEMP Out Of
Range
The temperature of the AG
module exceeds the
specified working
temperature range.
High
Make sure the AG
module is used
within the specified
temperature range.
Sample Line
Occluded
The sampling tube is
occluded.
Medium
Replace the
sampling tube.
AG No Adapter
No adapter is connected. Medium
Connect the adapter
correctly.
No Sample Line
No sampling tube is
connected.
Medium
Connect the
sampling tube
correctly.
AG Occlusion
AG module sample line
occluded
High
Replace the
sampling line.
Check
Watertrap/Sample
Line
Watertrap or sample line
falls off.
Low
1) Check whether
water trap is
installed normally.
2) Check whether
sample line is
installed normally.
AG Change
Watertrap
Malfunction in watertrap Medium
Replace the
watertrap.
Watertrap will be
full
Watertrap will be full. Medium
Replace the
watertrap.