Vital Signs Monitor User Manual Alarm Information
- 59 -
Message Cause Alarm Level Action Taken
NIBP Range Exceeded
Maybe the patient blood
pressure value is beyond
the measurement range.
High
Use other methods
to measure blood
pressure.
NIBP Loose Cuff
Cuff is not properly
wrapped or no cuff is
connected.
Low
Properly wrap the
cuff.
NIBP Interference
Signal noise is too large
or pulse rate is not regular
due to the patient
movement.
Low
Make sure that the
patient under
monitoring is
motionless.
T2A TEMP
Temp Comm Fail
TEMP module failure or
communication failure.
High (monitor
mode)
Low
(spot-checking
or Ward round
mode)
Stop using
measuring function
of TEMP module;
notify biomedical
engineer or
Manufacturer’s
service staff.
Temp Exceed Limit
The TEMP value is
beyond the range of 35.5
°C ~ 42 °C
Medium
Put the sensor into
the sensor bracket,
take it out and
measure again.
No TEMP Sensor
TEMP sensor is not
connected to the TEMP
module.
Low
Connect the sensor
and the monitor
well, and measure
again.
Ambient TEMP Too
High
The Sensor temperature is
higher than +40 °C
Low
Put the sensor into
the sensor bracket,
measure again after
the ambient
temperature reaches
normal value.
Ambient TEMP Too
Low
The Sensor temperature is
lower than +10 °C
Low