Vital Signs Monitor User Manual Alarm Information
- 60 -
Message Cause Alarm Level Action Taken
Probe Is Missing
The probe connected
cannot be identified.
Medium Please change probe
or contact
Manufacturer’s
service staff.
Probe TEMP Too High
The original temperature
of sensor >+33 °C and
+40 °C.
Low
Please wait for
probe temp drop
and start measuring.
Probe Heater Error
Heater error. Medium Stop using
measuring function
of TEMP module;
notify biomedical
engineer or
Manufacturer’s
service staff.
Temp SENSOR OFF
The probe is disconnected
from the patient.
Low
Connect the sensor
and the patient well,
and measure again.
Measure Site Error
The probe in use is not
consistent with the
measure position set on
the monitor.
Medium
Correctly set the
measure position on
the monitor.
TH TEMP
Temp exceed limit
The TEMP value is
beyond the range of 34 °C
~ 42.2 °C
Medium
Check the integrity
of the probe cover,
make sure it is
clean, and take a
new measurement.