Vital Signs Monitor User Manual Alarm Information
- 61 -
Message Cause Alarm Level Action Taken
F3000 TEMP
TEMP Error E1
System error during
synchronization.
Medium
Stop using
measuring function
of TEMP module;
notify biomedical
engineer or
Manufacturer’s
service staff.
TEMP Error E2
System error during
patient thermistor
calibration.
Medium
TEMP Error E3
System error during heater
thermistor calibration.
Medium
TEMP Error E4
System timing error.
Medium
Probe Heater error
Heater error.
Medium
TEMP Error P2
Monitor Mode patient
thermistor unstable or out
of range.
Low
TEMP Error P3
Monitor Mode heater
thermistor unstable or out
of range.
Low
TEMP Error P4
Predict Mode patient
thermistor unstable or out
of range.
Low
TEMP Error P5
Predict Mode heater
thermistor unstable or out
of range.
Low