Vital Signs Monitor User Manual Alarm Information
- 62 -
Message Cause Alarm Level Action Taken
TEMP Error P6
Unable to pre-heat probe
tip.
Low
Stop using
measuring function
of TEMP module;
notify biomedical
engineer or
Manufacturer’s
service staff.
NOTE: Measure
readings
displayed on the
screen are
unreliable when
the monitor
indicates Temp
Error P06.
Temp COMM Fail
TEMP module failure or
communication failure.
High (monitor
mode)
Low
(spot-checking
or Ward round
mode)
Stop using
measuring function
of TEMP module;
notify biomedical
engineer or
Manufacturer’s
service staff.
Temp exceed limit
The TEMP value is out of
the range of +30 °C ~ +43
°C.
Medium
Put the probe into
the probe well; take
it out and measure
again.
No TEMP Sensor
Probe configuration (or no
probe connected) error.
Low
Well connect the
probe and the
monitor, and
measure again.
Measure Site Error
The probe in use is not
consistent with the
measure position set on
the monitor.
Medium
Correctly set the
measure position on
the monitor.
Probe TEMP Too High
The original temperature
of sensor >+33 °C and
+40 °C.
Low
Please wait for
probe temp drop
and start measuring.