Vital Signs Monitor User Manual Alarm Information
- 63 -
Message Cause Alarm Level Action Taken
Others
Battery Low
Battery Low
High (monitor
mode)
Low
(spot-checking
or Ward round
mode)
Please change the
battery or charging.
Battery Error
Malfunction in Battery
Low
Replace the battery
and restart the
monitor. If the
problem persists,
notify the
manufacturer’s
service staff.
Recorder Out Of Paper
Recorder Out Of Paper Low
Please install the
paper.
Recorder Setup
Needed
The user presses the
RECORD button when
Recorder is not
configured.
Low
Notify the
manufacturer’s
service staff to
install and set the
recorder.
Insufficient storage
space
Less than 10M space is
left in the device.
Low
Delete some data in
the device or use
another device.
Read-only storage
device
The device is read-only. Low
Repair the device or
replace it with a
new one.