EasyManua.ls Logo

EDAN iM8 - Page 187

EDAN iM8
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User Manual Specifications
- 178 -
Warm-up
time
Display waveform within 20 s; Reach the design accuracy within 2 minutes.
Rise time
< 400 ms (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
< 500 ms (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Response
time
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Work mode
Standby, measure. Default: Standby
O
2
compensation
Range: 0% to 100%
Resolution: 1%
Default: 16%
N
2
O
compensation
Range: 0% to 100%
Resolution: 1%
Default: 0%
AG
compensation
Range: 0% to 20%
Resolution: 0.1%
Default: 0%
Humidity
compensation
method
ATPD (default), BTPS
Barometric
pressure
compensation
Automatic (Barometric pressure change will not bring additional deviation to
measuring accuracy).
Calibration
Support
Alarm
EtCO
2
, FiCO
2
, AwRR
Apnea
alarm delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Data Sample
Rate
100 Hz
EtCO
2
Change
1
AwRR 80rpm, meet the accuracy
mentioned above;
AwRR > 80 rpm, EtCO
2
descends
8%;
AwRR > 120 rpm, EtCO
2
descends
10%
with 2 m gas sampling tube, sample gas
flowrate: 100 ml/min)

Table of Contents

Related product manuals