Patient Monitor User Manual Alarm Information
- 55 -
Chapter 7 Alarm Information
7.1 Physiological Alarm Information
WARNING
During monitoring, the physiological alarms including ASYSTOLE, VFIB/VTAC, RESP
APNEA, SpO
2
No Pulse, CO
2
APNEA are preset to be on and cannot be turned off.
Message
Cause
Alarm Level
RR HIGH
RESP measuring value is above upper alarm
limit.
User-selectable
RR LOW
RESP measuring value is below lower alarm
limit.
User-selectable
RESP APNEA
RESP can not be measured within the set
apnea alarm delay time.
High
SpO
2
HIGH
SpO
2
measuring value is above upper alarm
limit.
User-selectable
SpO
2
LOW
SpO
2
measuring value is below lower alarm
limit.
User-selectable
PR HIGH
PR measuring value is above upper alarm
limit.
User-selectable
PR LOW
PR measuring value is below lower alarm
limit.
User-selectable
SpO
2
No Pulse
The signal of the measurement site is too
weak due to insufficient blood supply and
environmental factors, so the monitor can’t
detect the pulse signal.
High
NS HIGH
NIBP SYS measuring value is above upper
alarm limit.
User-selectable
NS LOW
NIBP SYS measuring value is below lower
alarm limit.
User-selectable
ND HIGH
NIBP DIA measuring value is above upper
alarm limit.
User-selectable
ND LOW
NIBP DIA measuring value is below lower
alarm limit.
User-selectable
NM HIGN
NIBP MAP measuring value is above upper
alarm limit.
User-selectable