Patient Monitor User Manual Default Settings
- 191 -
A3.4 RESP
RESP Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit
30
30
100
Alarm Low Limit
8
8
30
Apnea Time
20 s
Calculation Type
Auto
Resp Type
Sweep
12.5 mm/s
Amplitude
2
A3.5 SpO
2
SpO
2
Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit
100
100
95
Alarm Low Limit
90
90
88
Pitch Tone
Off
Sweep
12.5 mm/s
A3.6 PR
PR Settings
ADU
PED
NEO
PR Source
SpO
2
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit
120
160
200
Alarm Low Limit
50
75
100
Pulse Volume
3
Alarm Source
HR